It was reported that upon interrogation an error message was received stating that the generator was disabled due to burst watchdog timeout, and all settings were found to be at 0ma.It was noted that the patient was not affected.The device history records of the generator were reviewed.The generator passed final quality and functional specifications prior to release.Previous investigation has determined that certain m106 generators have a firmware that can inappropriately lead to burst watchdog timeout when the magnet output current is less than or equal to the autostimulation output current and the magnet is repeatedly swiped during an autostimulation burst.This event is consistent with this determination.The patient's settings were updated to prevent this issue.No further relevant information has been received to date.
|