WARSAW ORTHOPEDICS CD HORIZON® SOLERA® SPINAL SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 6440530
Device Problems Difficult to Remove (1528); Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/26/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information received from healthcare provider via manufacturer representative regarding an event happened during intra-op for a patient undergoing unknown spinal therapy.It was reported that, during the surgery when using the tool for detaching the screw to detach the screw tail cap, under the premise of correct operation, the screw tail cap did not detached, which caused the screw to slip, which led to the failure of normal implantation and use, resulting in the screw being scrapped, so applied for replacement.There was no health damage in patient was reported as a result of this event.There was no additional treatment or surgery performed.No further complications reported.2021-sep-03: additional information received.It was slipped during the implantation process, there was contact with the patient.There was no time delay in procedure reported as a result of this event.The surgery was successfully completed with another screw.No further complications reported.
 
Manufacturer Narrative
H3: product analysis of part#6440530 ; lot# 0841504w analysis summary: visual and macroscopic inspection confirmed the threads of the break off screw have been damaged.The thread crest and flank damage appear to have initiated at the start of the thread and is consistent around the damaged portion of the thread.Functional evaluation with a sample mas found the set screw unable to be fully engaged in the mas head.The above observations are consistent with misalignment of the mas and set screw threads during construct assembly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® SOLERA® SPINAL SYSTEM
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12504092
MDR Text Key272487659
Report Number1030489-2021-01182
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K143375
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6440530
Device Catalogue Number6440530
Device Lot Number0841504W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexMale
-
-