Catalog Number R5C4479C |
Device Problems
Fluid/Blood Leak (1250); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a leak was observed on the floor during use of a homechoice cassette.This was observed after completing peritoneal dialysis (pd) therapy.It was further reported that the leak was located closer toward the machine and "the clamps on the cassette did not make click sounds when secured".There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the actual device was discarded; however, a companion sample was received and evaluated.The companion sample was received unused and inside a closed pouch.Visual inspection with the naked eye did not identify any abnormalities that could have contributed to the reported condition.Homechoice machine testing was performed with no issues noted.The reported condition was not verified during the evaluation of the companion sample.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|