• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number L331
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/15/2020
Event Type  malfunction  
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Investigation of the available information determined this device exhibited oversensing of noise generated by the minute ventilation (mv) sensor that is related to a high impedance condition.Please see the description for more information regarding the specific circumstances of this event.This device was included in the minute ventilation sensor signal oversensing advisory population.
 
Event Description
It was reported that during a review of battery data for this pacemaker, two signal artifact monitor (sam) episodes were found, stored in (b)(6) 2020.In the episodes, minute ventilation (mv) sensor noise/artifact was observed on the non-boston scientific right atrial (ra) and right ventricular (rv) leads, which caused the mv sensor to disable.A review of the lead trends confirmed stable pacing impedance values over the past year.Currently, the mv sensor remains disabled, resolving the mv sensor noise issue.Additionally, there were two episodes stored showing noise oversensing on the ra and rv channels consistent with electromagnetic interference (emi), one in (b)(6) 2020 and another more recently in (b)(6) 2021.No adverse patient effects were reported.The device and leads remain implanted and in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI EL DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12515742
MDR Text Key272860729
Report Number2124215-2021-24671
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526559266
UDI-Public00802526559266
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/03/2017
Device Model NumberL331
Device Catalogue NumberL331
Device Lot Number700298
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/03/2015
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age39 YR
-
-