This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Investigation of the available information determined this device exhibited oversensing of noise generated by the minute ventilation (mv) sensor that is related to a high impedance condition.Please see the description for more information regarding the specific circumstances of this event.This device was included in the minute ventilation sensor signal oversensing advisory population.
|
It was reported that during a review of battery data for this pacemaker, two signal artifact monitor (sam) episodes were found, stored in (b)(6) 2020.In the episodes, minute ventilation (mv) sensor noise/artifact was observed on the non-boston scientific right atrial (ra) and right ventricular (rv) leads, which caused the mv sensor to disable.A review of the lead trends confirmed stable pacing impedance values over the past year.Currently, the mv sensor remains disabled, resolving the mv sensor noise issue.Additionally, there were two episodes stored showing noise oversensing on the ra and rv channels consistent with electromagnetic interference (emi), one in (b)(6) 2020 and another more recently in (b)(6) 2021.No adverse patient effects were reported.The device and leads remain implanted and in service.
|