• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIER; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Model Number IPN915189
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
Hemolok endo applier 5mm ml, clips not locking due to non functioning of jaws, jaws dislocated.
 
Event Description
Hemolok endo applier 5mm ml, clips not locking due to non functioning of jaws, jaws dislocated.
 
Manufacturer Narrative
(b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet , inc.Kenosha wi facility as part of a (b)(4)-piece lot in march of 2018.Evaluation of the returned instrument as received shows that the luer flush port cap is missing and the jaws are slightly loose and misaligned in the closed position thus we are able to validate this complaint.Further evaluation shows that the knob rotation and handle to jaw mechanisms are both dry and sluggish feeling suggesting that this instrument is in need of proper lubrication.After the initial evaluation this instrument was disassembled in order to evaluate its internal components and it was found that the inner drive rod (n00185) bosses are both damaged where they engage the jaws.(pie's attached) we suspect that the damaged drive rod bosses caused the jaws to become slightly loose and misaligned in the closed position.Mishandling of this device at the end user's facility is suspected.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL ML 5MM ENDO APPLIER
Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key12519210
MDR Text Key272904972
Report Number3011137372-2021-00262
Device Sequence Number1
Product Code GDO
UDI-Device Identifier24026704696813
UDI-Public24026704696813
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN915189
Device Catalogue Number544965
Device Lot Number06K1758656
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2021
Date Manufacturer Received10/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-