Model Number 71440472
Device Problems Failure to Align (2522); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/31/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that, during total knee arthroplasty, a genesis ii tibial tray implant impactor failed to grip onto the baseplate.In addition, on impaction the screw came loose.Surgery was resumed, after a non-significant delay, with a change in surgical technique.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII TIB TRY IMPL IMPACTOR
Type of Device
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12532311
MDR Text Key273316778
Report Number1020279-2021-07170
Device Sequence Number1
Product Code FSM
UDI-Device Identifier03596010329875
UDI-Public03596010329875
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440472
Device Catalogue Number71440472
Was Device Available for Evaluation? No
Date Manufacturer Received10/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-