H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one encor biopsy probe and vacuum assembly were returned for evaluation.During visual evaluation, the biopsy probe appeared to be clean and the aperture was received completely closed.Foreign material was noted on the device.Upon functional testing, the probe was loaded into the driver and calibrated successfully.The probe was calibrated an additional two times, both times were successful.No unusual noises were heard.Upon microscopic observation, the device was examined further under the microscope and residue was found to be on the aperture and cutter of the probe.Therefore, the investigation is confirmed for the reported foreign material issue as the foreign material was noted on the device.A definitive root cause for the reported foreign material issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: (expiry date: 09/2021).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: see h10.
|