• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. DISP FIRSTPASS STR PASSR SELF; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. DISP FIRSTPASS STR PASSR SELF; PASSER Back to Search Results
Catalog Number 22-4038
Device Problems Break (1069); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that during a shoulder arthroscopy, the jaw of the firstpass suture passer jammed and snapped when it bit through the tissue.The jaw was removed with arthroscopic graspers.The procedure was completed with non-significant delay using a back-up device.No other complications were reported.
 
Manufacturer Narrative
H2: h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A clinical review states this case reports the broken jaw of the fristpass suture passer was removed from inside of the patient with arthroscopic graspers.Based on the limited information provided the root cause of the reported jaw breakage could not be determined.According to report the procedure was completed with a non-significant delay using a back-up device.Since there was no patient injury or other complications reported, no further clinical/medical assessment is warranted at this time.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISP FIRSTPASS STR PASSR SELF
Type of Device
PASSER
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key12533868
MDR Text Key273379931
Report Number3006524618-2021-00938
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier00885556724545
UDI-Public885556724545
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 11/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number22-4038
Device Lot Number2073337
Was Device Available for Evaluation? No
Date Manufacturer Received11/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-