• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SYRINGE IRRIGATION BULK / CLEAN, 50ML; BULB SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SYRINGE IRRIGATION BULK / CLEAN, 50ML; BULB SYRINGE Back to Search Results
Model Number 0935280
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/30/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the bulb syringe was found to have a hair in it.
 
Event Description
It was reported that the bulb syringe was found to have a hair in it.
 
Manufacturer Narrative
The reported event was confirmed cause unknown.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.Visual evaluation of the returned sample noted one opened (without original packaging), used irrigation bulb syringe.Visual inspection of the sample noted that there was a strand of hair between the barrel and the bulb on the return sample.This does not meet the specification "no hair is permitted on the product.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labelling could not have prevented the reported failure.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE IRRIGATION BULK / CLEAN, 50ML
Type of Device
BULB SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12541068
MDR Text Key273609114
Report Number1018233-2021-06039
Device Sequence Number1
Product Code KYZ
UDI-Device Identifier10801741051378
UDI-Public(01)10801741051378
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number0935280
Device Catalogue Number0935280
Device Lot NumberNGEX0045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-