It was reported that there was an issue with pl569t-ligating clips m/l 12/box.According to the complaint description, it was repoted that they had a problem during an operation with a clip, the clip came off during the operation there was no described patient harm.Additional information was not provided nor available / was not available.Additional patient information is not available.The adverse event / malfunction is filed under aag reference (b)(4).Involved components: pl536r-shaft compl.D:10mm l:370mm-unknwon.Pl510r-handle f/pl506r & pl508r-unknown.
|
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: due to the fact that no lot number was provided, a review of the device history records for the complained device is not possible.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 3(5) x probability of occurrence(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
|