The device was returned for investigation.Upon evaluation of the device, the lamp heat-sink, lamp shaft, sink spring, and front pole were burned.In addition, a non-olympus lamp life meter reading over 500+hours, light output measured out of range, and worn out socket slider switch caused intermittent use of high intensity mode were note.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the event of a blurry image and the lamp heat sink, lamp shaft, sink spring, front pole being burned could not be identified.Olympus will continue to monitor field performance for this device.
|