Following notification of the reported event, steris quality requested that the cup subject of the event be returned for evaluation.Upon receipt, three cups were returned to steris (1 cup from lot number hc3450w2 and 2 cups from lot number hc2810r9).The returned cups were evaluated by steris and no issues with the welds for the lid were identified.No leaking was observed.Steris performed retain testing and confirmed there was no evidence of damage or leaking cups.Additionally, the device history records were reviewed and confirmed the lots were manufactured to specification; no abnormalities were identified.There have been no other complaints associated with these lots.No additional issues have been reported.
|