MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT XL VASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT XL VASCULAR STENT Back to Search Results
Catalog Number EX061503C
Device Problems Premature Activation (1484); Self-Activation or Keying (1557)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/08/2021
Event Type  malfunction  
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the lifestent vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 10/2022).
 
Event Description
It was reported that upon opening the outer package, the stent was allegedly found to be released.The procedure was completed using another device.There was no patient contact.
 
Event Description
It was reported that upon opening the outer package, the stent was allegedly found to be released.The procedure was completed using another device.There was no patient contact.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent are identified in d2 and g4.H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned unpacked and without original packaging.The deployment mechanism was found in unused condition, and the shipping lock was found correctly attached to the system.The stent was found 3mm partially deployed without tip to sheath gap which leads to confirmed result for premature deployment.Based on the information available a definite root cause could not be identified.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use state: 'examine the stent and delivery system device for any damage.If it is suspected that the sterility or performance of the device has been compromised, the device should not be used.Visually inspect the distal end of the delivery system catheter to ensure that the stent is contained within the sheath.Do not use if the stent is partially deployed.' h10: d4 (expiry date: 10/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Manufacturer Narrative
H10: the file was reassessed for reportability and determined to be no longer reportable.Since an initial mdr was submitted, therefore, the file will remain assessed as a malfunction.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent are identified in d2 and g4.H10: b5, d4 (expiry date: 10/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that prior to a stent placement procedure, the stent was allegedly found to be released upon opening the outer package.The procedure was completed using another device.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT XL VASCULAR STENT
Type of Device
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12576815
MDR Text Key274749307
Report Number9681442-2021-00542
Device Sequence Number1
Product Code NIP
UDI-Device Identifier04049519001487
UDI-Public(01)04049519001487
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P070014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEX061503C
Device Lot NumberANEX3166
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/27/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
Patient SexMale
Patient Weight60 KG
-
-