Catalog Number EX061503C |
Device Problems
Premature Activation (1484); Self-Activation or Keying (1557)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The catalog number identified has not been cleared in the us but is similar to the lifestent vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 10/2022).
|
|
Event Description
|
It was reported that upon opening the outer package, the stent was allegedly found to be released.The procedure was completed using another device.There was no patient contact.
|
|
Event Description
|
It was reported that upon opening the outer package, the stent was allegedly found to be released.The procedure was completed using another device.There was no patient contact.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent are identified in d2 and g4.H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned unpacked and without original packaging.The deployment mechanism was found in unused condition, and the shipping lock was found correctly attached to the system.The stent was found 3mm partially deployed without tip to sheath gap which leads to confirmed result for premature deployment.Based on the information available a definite root cause could not be identified.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use state: 'examine the stent and delivery system device for any damage.If it is suspected that the sterility or performance of the device has been compromised, the device should not be used.Visually inspect the distal end of the delivery system catheter to ensure that the stent is contained within the sheath.Do not use if the stent is partially deployed.' h10: d4 (expiry date: 10/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Manufacturer Narrative
|
H10: the file was reassessed for reportability and determined to be no longer reportable.Since an initial mdr was submitted, therefore, the file will remain assessed as a malfunction.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent are identified in d2 and g4.H10: b5, d4 (expiry date: 10/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that prior to a stent placement procedure, the stent was allegedly found to be released upon opening the outer package.The procedure was completed using another device.There was no patient contact.
|
|