Model Number MC2-5090S
Device Problem Insufficient Information (3190)
Patient Problem Failure of Implant (1924)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event Description
Information was received that the rod failed in the patient.No patient adverse event was reported.
 
Manufacturer Narrative
Even though no product has been returned, fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs.Was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
 
Event Description
Information was received that the rod's end cap disengaged.No patient adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key12584561
MDR Text Key274993741
Report Number3006179046-2021-00476
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00887517834041
UDI-Public887517834041
Combination Product (y/n)N
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMC2-5090S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-