STRYKER TRAUMA KIEL TIBIAL NAIL, STANDARD T2 TIBIA Ø11X300 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1822-1130S
Device Problems Separation Failure (2547); Misassembly by Users (3133)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/16/2021
Event Type  malfunction  
Manufacturer Narrative
Upon completion of the investigation any additional information will be communicated in a supplemental report.
 
Event Description
It was reported the nail was attached to the handle backwards and would not come off.
 
Event Description
It was reported the nail was attached to the handle backwards and would not come off.
 
Manufacturer Narrative
The reported event could be not confirmed, since the product was not returned for evaluation.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.The information for use states that ¿careful inspection and functional test of the device before use is the best method of determining the end of serviceable life for the medical device.Do not hit instruments unless they are specifically intended for impaction.Never hit targeting devices or elastosil handles other than those with an integrated strike plate! always treat the instrument carefully to avoid surface damage or alterations to the geometry.¿ no indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.H3 other text : device disposition is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL NAIL, STANDARD T2 TIBIA Ø11X300 MM
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
sharon rivas
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key12590478
MDR Text Key275180616
Report Number0009610622-2021-00742
Device Sequence Number1
Product Code HSB
UDI-Device Identifier04546540198013
UDI-Public04546540198013
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model Number1822-1130S
Device Catalogue Number18221130S
Device Lot NumberK0B9164
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/23/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/08/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-