Model Number EG29-I10C |
Device Problems
Loss of or Failure to Bond (1068); No Display/Image (1183); Fluid/Blood Leak (1250); Moisture Damage (1405); Use of Device Problem (1670); Optical Distortion (3000)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/16/2021 |
Event Type
malfunction
|
Event Description
|
The images fades over, the electrical safety test failed, there is moisture inside the cmos chip.This event occurred at the time of before use.There was no report of patient harm.
|
|
Manufacturer Narrative
|
This device is classified as import for export, therefore 510k is not applicable.Model eg29-i10c-us is available in the usa with a 510k number k190805.If additional information becomes available, a supplemental report will be filed with the new information.
|
|
Manufacturer Narrative
|
Evaluation summary: the cmos chip and the distal body replaced.
|
|