Was reported that the patient was seen in clinic and had their device checked for the first time since (b)(6) 2019.The doctor noted that the patient's normal mode and magnet mode output currents were both set to 0 ma although the other settings were intact.The neurologist was able to correct the settings.The family reported that the patient's seizures had worsened since 2020.It is known that m102 generator that the patient is implanted with is susceptible to settings changes if diagnostics tests are interrupted (this happens because on m102 generators, diagnostics tests reprogram the patient's settings to run the diagnostic test and then will reprogram the generator, back to intended settings, so if the diagnostics test is interrupted, then settings may not be corrected to intended settings).If a final interrogation prior to ending a programming session to check settings is not performed as recommended, the patient may be left at this settings change.The patient's output current being disabled is consistent with this interrupted diagnostics.The doctor reported that there was no mris (which would require disablement) since the visit.No further relevant information has been received to date.
|
The increased seizures were further described as a marked decline in seizure control with up to 1-3 seizures daily, frequent convulsions, and falls, resulting in emergency room visits and a need to adjust aeds with the addition of a new anti-seizure drug.The patient also moved in with their parents for safety.No further relevant information has been received to date.
|