The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: the returned unit passed all specific functional testing requirements except for the lock having rotational movement when the unit is not under pressure.This would not have caused a slippage.When the unit is properly positioned and put under pressure, it would not have slipped.All worn components were replaced and general maintenance was performed.Root cause: evaluation found no device deficiencies that would have contributed to the reported complaint.Repairs could not duplicate slippage.Probable root cause for reported complaint is improper placement of the skull clamp.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|