Model Number 101/860/075CZ
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2021
Event Type  malfunction  
Event Description
Information was received by smiths medical that the patient moved, and it caused removal of the product to occur.The product had been in use from aug 10th to sept.5.No adverse affects to patient reported.
 
Manufacturer Narrative
Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#(b)(4).One decontaminated sample was returned for evaluation.Under visual inspection, it was noticed that inflation line was detached from pilot balloon.Tear in cuff was observed therefore detached pilot balloon is considered to be the source of leak.Thus confirming the reported failure mode.No other analysis was performed.Root cause was found to be: not enough tetrahydrofuran solvent at joint.Corrective actions will be addressed and implemented trough a non-conformance report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES BLUSELECT
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
akasaka, minato-ku, tokyo
minneapolis, MN 55442
MDR Report Key12601861
MDR Text Key275886086
Report Number3012307300-2021-09865
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076141
UDI-Public15019517076141
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/075CZ
Device Catalogue Number101/860/075CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/24/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-