OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to omsc for evaluation but was returned to (b)(4).(b)(4) checked the subject device and found the reported phenomenon.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at the service department of olympus (b)(4), it was found that there were a lot of oily corrosions on the main board, the electropneumatic proportional valve, the manifold unit and the tubes inside the subject device.The subject device had been returned to (b)(4) because the front panel had damage.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The cause of this phenomenon could not be identified because the product was not returned.The corrosion may be attributed to usage and storage environment.The phenomenon seems to have occurred due to the user¿s handling.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12602614
MDR Text Key281508783
Report Number8010047-2021-12948
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-