Model Number 8145 |
Device Problems
Failure to Conduct (1114); Impedance Problem (2950)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/17/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
If device is received, an evaluation is expected.
|
|
Event Description
|
It was reported that the patient was unable to initiate stimulation.Reportedly the stimulation lead electrodes were not functioning.Patient's surgeon decided to revise the complete reactiv8 system (stimulation leads, and implantable pulse generator).During revision surgery, the lead was damaged resulting in the distal three electrodes being unable to be explanted.The replacement system was successfully implanted to resolve the issue.
|
|
Manufacturer Narrative
|
Functional testing was performed on the returned implantable pulse generator (ipg).Ipg was found to be operating as intended.Visual inspection of the leads was performed due to device damage during explant.No fractures were observed on the first lead; functional testing was unable to be performed due to lead being cut during explant at the proximal end.Three fractured wires were observed on the second lead 390 millimeters from the proximal end.It is likely that the fractures were present for an extended period prior to explant because of the observed smoothness of the wire ends.The root cause of the reported out of range issue is the observed fractures.
|
|
Search Alerts/Recalls
|