Model Number H7493941828250 |
Device Problems
Device Dislodged or Dislocated (2923); Material Deformation (2976); Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/05/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that stent moved on balloon was encountered.A 2.50 x 28mm synergy xd drug eluting stent was selected for preparation.However there was bend right behind the balloon and the stent moved on the balloon.The procedure was completed successfully.The device was not used for this procedure.There were no patient complications reported.
|
|
Event Description
|
It was reported that stent moved on balloon was encountered.A 2.50 x 28mm synergy xd drug eluting stent was selected for preparation.However there was bend right behind the balloon and the stent moved on the balloon.The procedure was completed successfully.The device was not used for this procedure.There were no patient complications reported.It was further reported that the stent was damaged and not fully on the balloon.
|
|
Event Description
|
It was reported that stent moved on balloon was encountered.A 2.50 x 28mm synergy xd drug eluting stent was selected for preparation.However there was bend right behind the balloon and the stent moved on the balloon.The procedure was completed successfully.The device was not used for this procedure.There were no patient complications reported.It was further reported that the stent was damaged and not fully on the balloon.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: synergy xd mr us 2.50 x 28mm stent delivery system (sds) was returned for analysis.Examination of the crimped stent via scope found stent damage.Stent struts on the proximal end were flared.The stent showed no signs of movement and was set between the proximal and distal markerbands.The undamaged crimped stent outer diameter was measured by snap gauge and the result was within max crimped stent profile measurement.The balloon cones were reviewed via scope, and no issues were noted.The balloon returned deflated and the proximal balloon cone appeared bunched.A visual and tactile examination of the hypotube found multiple kinks.Examination of the tip via scope found damage to the tip.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
|
Search Alerts/Recalls
|