Catalog Number 07464215190 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is ongoing.
|
|
Event Description
|
The initial reporter received a questionable elecsys vitamin d assay result for one patient tested on a cobas e411 disk.On (b)(6) 2021, the patient's sample had an initial result of >100ng /ml.This result was reported outside of the laboratory.On (b)(6) 2021, the result was questioned by the physician prompting a rerun of the sample.The same sample was then rerun on the same instrument.The repeat result was again >100 ng/ml.The same sample was then sent to another laboratory.The repeat result from this laboratory was 10.1 ng/ml.The serial number of the instrument is (b)(4).
|
|
Manufacturer Narrative
|
It was discovered that duplicate cases were created for the same event.Any additional information will be filed as a supplemental to mdr 1823260-2021-02895.
|
|