• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS VITAMIN D TOTAL II; VITAMIN D TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS VITAMIN D TOTAL II; VITAMIN D TEST SYSTEM Back to Search Results
Catalog Number 07464215190
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/04/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.
 
Event Description
The initial reporter received a questionable elecsys vitamin d assay result for one patient tested on a cobas e411 disk.On (b)(6) 2021, the patient's sample had an initial result of >100ng /ml.This result was reported outside of the laboratory.On (b)(6) 2021, the result was questioned by the physician prompting a rerun of the sample.The same sample was then rerun on the same instrument.The repeat result was again >100 ng/ml.The same sample was then sent to another laboratory.The repeat result from this laboratory was 10.1 ng/ml.The serial number of the instrument is (b)(4).
 
Manufacturer Narrative
It was discovered that duplicate cases were created for the same event.Any additional information will be filed as a supplemental to mdr 1823260-2021-02895.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS VITAMIN D TOTAL II
Type of Device
VITAMIN D TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key12627952
MDR Text Key278555405
Report Number1823260-2021-02968
Device Sequence Number1
Product Code MRG
Combination Product (y/n)N
Reporter Country CodeTC
PMA/PMN Number
K113546
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Catalogue Number07464215190
Device Lot Number547353
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-