Catalog Number V041183000000 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event it was reported that a vdw.Connect drive motor in combination with a vdw.Connect locate gave incorrect measurements.No injury occured.
|
|
Manufacturer Narrative
|
Handpiece (b)(6) (measurement socket/charging socket broken) defect, replaced with sn10616.Measurement cable for combined measuring (plug bent) defect, replaced.Contra angle (b)(6) (2015) (head drive worn) defect replaced with (b)(6).Locate (b)(6) (2019) without error.
|
|
Search Alerts/Recalls
|