Catalog Number A1059
Device Problems Break (1069); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the swivel lock on the mayfield modified skull clamp (a1059) was broken, ratchet mechanism was sticking and does not open or close easily.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
 
Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h6, h10.The mayfield skull clamp (a1059) was returned for evaluation: failure analysis - the investigation of the returned device verified the reported complaint.Unit received with the lock having rotational and lateral movement and a residue buildup was present.New components was added to replace worn internal parts along with general maintenance and cleaning.Root cause - based on the evaluation by integra service and repair, the root cause is most likely normal wear over time, in addition to the need for routine preventive maintenance.The unit is 11 years old, well beyond integra's recommended 7 year life cycle (manufactured 2010) no further investigation required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.
 
Event Description
N/a.
 
Manufacturer Narrative
The mayfield modified skull clamp (a1059) was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause(s) of the reported issue could not be determined.The reported condition is likely caused by improperly handling/ wear and tear of the device.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12635981
MDR Text Key276590462
Report Number3004608878-2021-00599
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received12/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown