Model Number AIA-2000 ST |
Device Problems
Volume Accuracy Problem (1675); Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error logs, but did not reproduce error.While priming the wash, fse noticed air pocket above the wash syringe of bf probes #1 and #2.Fse replaced the 3way valves to prevent air from going into the tubing.Fse cleaned and inspected both wash syringes, ran daily maintenance and quality control without any errors and within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-2000 operator's manual under appendix 4: error messages states the following: [2239] b/f probe 1 purge failure.Cause : the overflow sensor failed to detect liquid even after the washer was purged.Solution : air may be trapped in the washer tubing.Purge any remaining air by performing the priming operation and check for the presence of air in the washer tubing.If retry fails, contact tosoh service center or local representatives.The most probable cause of the reported event is due to failure of the 3 way valves of the washer.
|
|
Event Description
|
A customer reported getting error message "2239 bf probe 1 purge failure" on the aia-2000 instrument.The customer stated all bf probes were reporting purge failures, while all tubing were connected.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for intact parathyroid hormone (ipth).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Search Alerts/Recalls
|
|
|