(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was returned to fisher & paykel healthcare (f&p) in new zealand where it was visually inspected and analysed.Results: visual inspection of the mr290v chamber confirmed that there were no notable damages.A leak was not identified when the device was connected to a water source.Then the device was pressure tested, and a leak was observed between the chamber dome and chmaber base as reported.After dis-assembling the chamber, it was found that the chamber dome was deformed.Conclusion: we are unable to determine the cause of the reported fault.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.The subject chamber would have met the required specification prior to distribution.Our user instructions that accompany the mr290v vented autofeed humidification chamber states the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." -"use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions, could lead to a loss of ventilation pressure.".
|