Model Number 26432020-1
Device Problem Overheating of Device (1437)
Patient Problem Burn(s) (1757)
Event Type  Injury  
Manufacturer Narrative
Refer to medwatch report #9610617-2021-00068.
 
Event Description
As per a manufacturer incident report we received from the factory in (b)(6): the "optitherm" accessory, which was intended for preheating the gas to be insufflated has overheated far beyond the intended alarm limit value of 41 °c.This situation was not detected by the main unit.This circumstance was not detected by the main unit, so no alarm was triggered.Due to the direct skin contact, a large burn corresponding to the size of the housing burn occurred.Two devices were involved (optitherm and thermoflator).This report is for device 2 (thermoflator).(mfr's.Internal complaint #(b)(4)).
 
Manufacturer Narrative
Since the devices were locked by the customer, a physical investigation could not be performed.Based on the investigation report (dated (b)(6) 2021), the most probable root cause is a defect of the heat exchanger.It is probable that the plug of the gas heater was twisted due to mechanical stress, which could have led to the reported failure.If a physical investigation is possible, a revised report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOFLATOR
Type of Device
INSUFFLATION DEVICE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM 
MDR Report Key12642608
MDR Text Key281209769
Report Number9610617-2021-00069
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number26432020-1
Device Catalogue Number26432020-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-