• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Catalog Number CLXECP
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/17/2021
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction tubing leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot j152 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot j152 shows no trends.Trends were reviewed for complaint category, tubing leak.No trends were detected for this complaint category.This assessment is based on the information available at the time of the investigation.At the time of this report, the analysis of the returned kit is still in progress.A supplemental report will be filed when the analysis is complete.(b)(4).
 
Event Description
The customer contacted mallinckrodt to report they experienced a tubing leak with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported they noticed a crack in the tubing and observed a blood leak from the collect line tubing.The customer aborted the ecp treatment and residual blood within the kit was not returned to the patient.The customer reported the patient was in stable condition.The customer returned the kit for investigation.
 
Manufacturer Narrative
The complaint kit and smart card were returned for investigation.Review of the data recorded on the smart card showed multiple alarm #52: collect line air detected alarms occurred during the treatment.The received kit was intact and configured for double needle mode.Examination of the needle-free port on the collect line showed that the blue insert in the middle of the port was oval rather than circular.There was some dried blood around the top of the needle-free port.The port was pressure tested and a leak was confirmed from the port.A material trace of the needle-free port and its components used to build lot j152 found no related non-conformances.A device history record review did not identify any related non-conformances and this kit lot had passed all lot release testing.The root cause of the reported tubing leak was most likely due to a manufacturing operator error.Manufacturing operators are to visually inspection all needle-free ports to ensure they are circular in shape and exhibit no visible damage.No further action is required at this time.This investigation is now complete.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
blanchardstown, dublin D15 T X2V
EI  D15 TX2V
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
shelbourne building
53 frontage rd suite 300
hampton, NJ 08827
MDR Report Key12642870
MDR Text Key279266500
Report Number2523595-2021-00087
Device Sequence Number1
Product Code LNR
Combination Product (y/n)Y
Reporter Country CodeUK
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/01/2022
Device Catalogue NumberCLXECP
Device Lot NumberJ152
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-