Model Number 74011857
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that a journey ii cr lkg fem imp bumper right is cracked.It remains unknown when this problem was noticed/occurred and if a patient was involved.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned jrny ii cr lkg fem imp bumper rt confirmed the device is broken in half and both pieces were returned.This device also has extensive wear usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II CR LKG FEM IMP BUMPER RT
Type of Device
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12643312
MDR Text Key276849507
Report Number1020279-2021-07593
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00885556085233
UDI-Public00885556085233
Combination Product (y/n)N
PMA/PMN Number
K121443
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/27/2017
Device Model Number74011857
Device Catalogue Number74011857
Device Lot Number16MBG0083
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2021
Date Manufacturer Received10/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-