MAUDE Adverse Event Report: BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM Back to Search Results
Catalog Number 256082
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Hypersensitivity/Allergic reaction (1907)
Event Date 09/16/2021
Event Type  Injury  
Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported while testing for sars cov-2 a patient experienced an allergic reaction.The patient went to the er as a result of the reaction.Eua#:(b)(4)."it was reported that patient experience allergic reaction to swab for cat 256082.".
 
Manufacturer Narrative
H6: investigation: this statement summarizes the investigation results regarding the complaint that alleges swab allergic reaction results when using kit rapid detection of sars-cov-2 veritor (material # 256082), batch number 1215732.Bd quality performs a systematic approach to investigate swab allergic reaction complaints.This approach involves review of manufacturing batch history records, testing of retention samples, and testing of customer returned samples, if applicable.An investigation was performed on the batch number provided and no relevant issue was found.The complaint was unable to be confirmed.The root cause could not be identified.A trend analysis for swab allergic reaction was conducted, no adverse trend was identified.Bd quality will continue to monitor for trends.There were no corrective actions taken at this time.
 
Event Description
It was reported while testing for sars cov-2 a patient experienced an allergic reaction.The patient went to the er as a result of the reaction.Eua#: eua (b)(4)."it was reported that patient experience allergic reaction to swab for cat 256082.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD RAPID DETECTION OF SARS-COV-2 VERITOR
Type of Device
CORONAVIRUS ANTIGEN DETECTION SYSTEM
Manufacturer (Section D)
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
9 rui pu road export zone b
suzhou
CH 
Manufacturer (Section G)
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
9 rui pu road export zone b
suzhou
CH  
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12643702
MDR Text Key276946438
Report Number3006948883-2021-00974
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/08/2021
Device Catalogue Number256082
Device Lot Number1215732
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/03/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-