H6: investigation: this statement summarizes the investigation results regarding the complaint that alleges swab allergic reaction results when using kit rapid detection of sars-cov-2 veritor (material # 256082), batch number 1215732.Bd quality performs a systematic approach to investigate swab allergic reaction complaints.This approach involves review of manufacturing batch history records, testing of retention samples, and testing of customer returned samples, if applicable.An investigation was performed on the batch number provided and no relevant issue was found.The complaint was unable to be confirmed.The root cause could not be identified.A trend analysis for swab allergic reaction was conducted, no adverse trend was identified.Bd quality will continue to monitor for trends.There were no corrective actions taken at this time.
|