• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT Back to Search Results
Model Number AVSM10040
Device Problems Obstruction of Flow (2423); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Aneurysm (1708); Restenosis (4576)
Event Date 05/24/2021
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 07/2022).Device not returned.
 
Event Description
It was reported through the results of a clinical trial, that approximately six months and fifteen days post index procedure, the subject developed 70% of target lesion stenosis.Standard percutaneous transluminal angioplasty (pta) was used to successfully treat the target lesion.The current patient status was not provided.
 
Manufacturer Narrative
H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the device was not returned for evaluation.The stent remains implanted.No x-ray images were provided for evaluation.Therefore, the alleged restenosis can not be reproduced.Based on the information available the investigation is inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, it was found that the instructions for use sufficiently address the potential risk.Based on the instructions for use, complications and adverse events associated with the use of the covera vascular covered stent may include the usual complications associated with endovascular stent and covered stent placement and dialysis shunt revisions, which includes restenosis of the target vessel.Regarding pre and post dilation the instructions for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated" and "post dilate the covered stent with an angioplasty balloon sized appropriately as to ensure complete wall apposition to the reference vessel.Avoid balloon dilation in the healthy, non-stenosed segment of the vessel." h10: d4 (expiry date: 07/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported through the results of a clinical trial, that approximately six months and fifteen days post index procedure, the subject developed 70% of target lesion stenosis.A standard percutaneous transluminal angioplasty (pta) was used to successfully treat the target lesion.The current patient status was not provided.
 
Manufacturer Narrative
H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the device was not returned for evaluation.The stent remains implanted.No x-ray images were provided for evaluation.Therefore, the alleged restenosis can not be reproduced.Pre dilation as well as post dilation was performed.Based on the information available the investigation is inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, it was found that the instructions for use sufficiently address the potential risk.Based on the instructions for use, complications and adverse events associated with the use of the covera vascular covered stent may include the usual complications associated with endovascular stent and covered stent placement and dialysis shunt revisions, which includes restenosis of the target vessel.Regarding pre and post dilation the instructions for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated" and "post dilate the covered stent with an angioplasty balloon sized appropriately as to ensure complete wall apposition to the reference vessel.Avoid balloon dilation in the healthy, non-stenosed segment of the vessel." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported through the results of a clinical trial, that approximately six months and fifteen days post index procedure, the subject developed 70% of target lesion stenosis.A standard percutaneous transluminal angioplasty (pta) was used to successfully treat the target lesion.The current patient status was not provided.
 
Manufacturer Narrative
H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the device was not returned for evaluation.The stent remains implanted.No x-ray images were provided for evaluation.Therefore, the alleged restenosis cannot be reproduced.Pre dilation as well as post dilation was performed.Based on the information available the investigation is inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, it was found that the instructions for use sufficiently address the potential risk.Based on the instructions for use, complications and adverse events associated with the use of the covera vascular covered stent may include the usual complications associated with endovascular stent and covered stent placement and dialysis shunt revisions, which includes restenosis of the target vessel.Regarding pre and post dilation the instructions for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated" and "post dilate the covered stent with an angioplasty balloon sized appropriately as to ensure complete wall apposition to the reference vessel.Avoid balloon dilation in the healthy, non-stenosed segment of the vessel." h10: d4 (expiry date: 07/2022), g3.H11: h6 (device).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported through the results of a clinical trial, that approximately six months and fifteen days post index procedure, the patient developed 70% of target lesion stenosis.Standard percutaneous transluminal angioplasty (pta) was used to successfully treat the target lesion.The current patient status was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA VASCULAR COVERED STENT
Type of Device
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12653252
MDR Text Key277131067
Report Number9681442-2021-00555
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741106385
UDI-Public(01)00801741106385
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Study,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAVSM10040
Device Catalogue NumberAVSM10040
Device Lot NumberANEU2369
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age83 YR
Patient SexFemale
Patient Weight75 KG
Patient EthnicityHispanic
Patient RaceWhite
-
-