MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SPYGLASS DS; LED LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SPYGLASS DS; LED LIGHT SOURCE Back to Search Results
Model Number M00546650
Device Problems Poor Quality Image (1408); Appropriate Term/Code Not Available (3191); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
Note: this report pertains to two spyscope ds ii catheter and a spyglass ds controller used during the same procedure.It was reported to boston scientific corporation that two spyscope ds ii catheter and a spyglass ds controller were used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed in the common bile duct on (b)(6) 2021.During the procedure, the controller kept turning on and off affecting image and usage of two spyscope ds ii catheter.There were purple lines came across the screen and the image freeze and delayed.The controller was rebooted, the spyscope ds ii catheter was unplugged multiple times and plugged back into the controller, wires and power cables were all checked; and it work for a few minutes then experience problems again.A second spyscope ds ii catheter was used and similar problems occurred after a few minutes.The procedure was not completed and will be rescheduled.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPYGLASS DS
Type of Device
LED LIGHT SOURCE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
ENERCON TECHNOLOGIES
25 northbrook dr
665
gray ME 04039
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key12665289
MDR Text Key277397345
Report Number3005099803-2021-05265
Device Sequence Number1
Product Code NTN
UDI-Device Identifier08714729874348
UDI-Public08714729874348
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142922
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00546650
Device Catalogue Number4665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/03/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-