Model Number 11996 |
Device Problem
Failure to Deliver Shock/Stimulation (1133)
|
Patient Problems
Burn(s) (1757); Ventricular Fibrillation (2130)
|
Event Date 10/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Physio-control contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
|
Event Description
|
The customer contacted physio-control to report that their device failed to deliver shock while in use with a patient.This issue is patient related; however there was no adverse event reported.
|
|
Manufacturer Narrative
|
Section h5 of initial mfr 0003015876-2021-02037, labeled for single use, states no.Section h5 has been updated to yes.The customer returned all unused electrodes of the same lot to stryker redmond for evaluation.All electrodes performed normally when tested, and no defects in adhesion or gel were detected.The electrodes were applied to the technician's skin for over 2 hours with no adhesion loss or redness when removed.Stryker was unable to replicate issue.No root cause could be determined.
|
|
Event Description
|
The customer contacted physio-control to report that their device failed to deliver shock while in use with a patient.This issue is patient related; however there was no adverse event reported.
|
|
Search Alerts/Recalls
|