• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-AVHD-DF16
Device Problems Material Frayed (1262); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/27/2021
Event Type  malfunction  
Manufacturer Narrative
One bi-directional, curve d-f, sensor enabled, advisor hd grid mapping catheter was received for evaluation.Visual inspection on the complaint device noted the distal coupler was fractured and detached.In addition, the nitinol frame and wires were torn.The catheter was inserted and withdrawn from a stock 8.5f sheath with no resistance or anomalies noted.The instructions for use states ¿insert the distal tip section of catheter into an 8.5 f minimum introducer (not included) using the insertion tool.¿ an 8.0f sheath was used during the procedure.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn wires and frame and the fractured and detached distal coupler is consistent with incorrect use of the device.
 
Event Description
This report is being submitted based on analysis which revealed exposed wires in the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12689461
MDR Text Key278141667
Report Number3005334138-2021-00657
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05415067028198
UDI-Public05415067028198
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Model NumberD-AVHD-DF16
Device Catalogue NumberD-AVHD-DF16
Device Lot Number7907748
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-