Model Number H700488
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2021
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported thermal event could not be determined.
 
Event Description
Following the procedure, an outlet cable that was plugged into an outlet was connected to the generator grounding port and the metal inside became burnt.There were no adverse consequences to the user who connected the outlet cable into the generator grounding port.
 
Manufacturer Narrative
Additional information: d9, g3, h2, h3, h6.One ampere¿ rf ablation generator was received into the lab for analysis.Visual inspection of the returned ampere generator revealed a damage/marred indifferent electrode connector pin which would have led to the reported complaint.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the reported event was confirmed and isolated to a damage/marred indifferent electrode connector pin.It is unknown how the damage occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPERE¿ RF ABLATION GENERATOR
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12704763
MDR Text Key278579515
Report Number2184149-2021-00338
Device Sequence Number1
Product Code LPB
UDI-Device Identifier05415067004888
UDI-Public05415067004888
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH700488
Device Catalogue NumberH700488
Device Lot Number4820865
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-