Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/13/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Visual examination of the returned product identified pitting and discoloration on the outer radius of the liner.The four corners of the locking feature at the apex of the liner show small nicks and burrs.No signs of fracture around the locking screw were observed.Deformation and gouging were observed around the rim of the inner radius.The locking screw rotates freely within the provisional.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that receiving found a small chunk broken out of the inner lip of the provisional g7 liner.No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 DM PROV LINER 46MM G
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12711005
MDR Text Key278764375
Report Number0001825034-2021-02960
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150522
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial
Report Date 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110024733
Device Lot Number674170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-