Catalog Number 31-424204 |
Device Problem
Fracture (1260)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 10/06/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported there was a foreign body left in patient after closing of hip.It was discovered in post op x-ray during patient recovery in pacu.The foreign body was believed to be piece of the flexible drill shaft.The surgeon attempted a second surgery later that day to retrieve the debris/foreign object but was unsuccessful.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: g3; h2; h3; h4; h6.Radiographs were provided and reviewed by a health care professional.Review of the available records identified anatomic alignment of the left hip arthroplasty.Retained metallic foreign object as noted device history record (dhr) was reviewed and no discrepancies were found.Root cause of the event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.
|
|
Event Description
|
No further event information available at the time of this report.
|
|