SIEMENS HEALTHCARE GMBH-AT ARTIS ZEE CEILING; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10094137
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2021
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zee ceiling system.The user reported that one cable roller wheel located in the dcs (display ceiling suspension) is defective.There is no report of impact to the state of health of any patient or user involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a user error.The investigation was performed considering complaint description, cs reports, system history, and system log files.It was reported that the cable trolley slipped out of the rail and fell down.The investigation revealed that a bearing seal on one of the three wheels of the cable trolley was broken.This kind of defect occurs when forces perpendicular to the movement direction are applied.If the wheel is already defective in this way, the cable trolley can slip out of the guide rail, as it performs slight oscillating when the cable trolley is pulled.The cable trolley cannot fall down completely as it is still attached to the wiring harness that hangs in loops from the ceiling.The investigation did not show a clear root cause for the external forces which damaged the wheel; but according to the system experts, the most likely cause is repeated extraordinary mechanical collisions with obstacles that could damage the cable trolley.This can result in a situation in which the wheel is broken from the mounting bracket.Siemens service resolved the error by replacing the cable trolley.The occurrence rate was checked and a possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE CEILING
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
6104486461
MDR Report Key12744990
MDR Text Key279887411
Report Number3004977335-2021-02507
Device Sequence Number1
Product Code OWB
UDI-Device Identifier04056869010052
UDI-Public04056869010052
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K181407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 11/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10094137
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/02/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-