Siemens has completed an investigation of the reported event.The root cause was determined to be a user error.The investigation was performed considering complaint description, cs reports, system history, and system log files.It was reported that the cable trolley slipped out of the rail and fell down.The investigation revealed that a bearing seal on one of the three wheels of the cable trolley was broken.This kind of defect occurs when forces perpendicular to the movement direction are applied.If the wheel is already defective in this way, the cable trolley can slip out of the guide rail, as it performs slight oscillating when the cable trolley is pulled.The cable trolley cannot fall down completely as it is still attached to the wiring harness that hangs in loops from the ceiling.The investigation did not show a clear root cause for the external forces which damaged the wheel; but according to the system experts, the most likely cause is repeated extraordinary mechanical collisions with obstacles that could damage the cable trolley.This can result in a situation in which the wheel is broken from the mounting bracket.Siemens service resolved the error by replacing the cable trolley.The occurrence rate was checked and a possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
|