• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS FEM LT SZ 5 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS FEM LT SZ 5 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL Back to Search Results
Model Number 1504-10-125
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Joint Laxity (4526)
Event Date 10/21/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Patient was revised due to instability, medial tibial collapse and loosening of the tibial component at the cement to implant interface.Competitor cement was used.The patella component was retained while all other components was revised.Doi: (b)(6) 2016 dor: (b)(6) 2021 left knee.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM LT SZ 5 NAR CEM
Type of Device
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12751968
MDR Text Key280140871
Report Number1818910-2021-24416
Device Sequence Number1
Product Code OIY
UDI-Device Identifier10603295041702
UDI-Public10603295041702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P830055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number1504-10-125
Device Catalogue Number150410125
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ATTUNE FB TIB BASE SZ 5 CEM; ATTUNE MEDIAL ANAT PAT 38MM; ATTUNE PS FB INSRT SZ 5 7MM; ATTUNE PS FEM LT SZ 5 NAR CEM; COMPETITOR CEMENT; ATTUNE FB TIB BASE SZ 5 CEM; ATTUNE MEDIAL ANAT PAT 38MM; ATTUNE PS FB INSRT SZ 5 7MM; ATTUNE PS FEM LT SZ 5 NAR CEM; COMPETITOR CEMENT
Patient Outcome(s) Required Intervention;
Patient Age66 YR
Patient SexFemale
Patient Weight127 KG
-
-