Model Number XS PST |
Device Problem
Circuit Failure (1089)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/14/2021 |
Event Type
malfunction
|
Event Description
|
We received an allegation of a display issue with a coaguchek xs meter.The customer reported that the meter was not working.She further stated that there was what looked like half of a number 1 on the screen.Upon completion of a display check, no segments were missing.The patient reported that three 8's kept appearing and disappearing from the screen and that they were unable to duplicate the half of a number 1 on the screen.The customer replaced the batteries but stated that the device was still not working.
|
|
Manufacturer Narrative
|
The meter was requested for investigation.Initial reporter occupation is patient/consumer.
|
|
Manufacturer Narrative
|
The meter was returned for investigation.The display did not show any noticeable contrast errors during the examination.The investigation shows that the battery compartment is contaminated by a leaked battery and the circuit board is contaminated by penetrated liquid which affects the conductive rubber contacts.The root cause is determined to be improper handling or maintenance by the customer.
|
|