• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Model Number XS PST
Device Problem Circuit Failure (1089)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2021
Event Type  malfunction  
Event Description
We received an allegation of a display issue with a coaguchek xs meter.The customer reported that the meter was not working.She further stated that there was what looked like half of a number 1 on the screen.Upon completion of a display check, no segments were missing.The patient reported that three 8's kept appearing and disappearing from the screen and that they were unable to duplicate the half of a number 1 on the screen.The customer replaced the batteries but stated that the device was still not working.
 
Manufacturer Narrative
The meter was requested for investigation.Initial reporter occupation is patient/consumer.
 
Manufacturer Narrative
The meter was returned for investigation.The display did not show any noticeable contrast errors during the examination.The investigation shows that the battery compartment is contaminated by a leaked battery and the circuit board is contaminated by penetrated liquid which affects the conductive rubber contacts.The root cause is determined to be improper handling or maintenance by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key12758429
MDR Text Key280523739
Report Number1823260-2021-03233
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702125100
UDI-Public00365702125100
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberXS PST
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received11/17/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-