INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD COMPOSITE SERIES SKULL CLAMP; ¿COMPOSITE SERIES¿ Back to Search Results
Model Number A3059
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Mayfield composite series skull clamp (a3059) was returned for evaluation.Device history record (dhr): the dhr was reviewed and show no abnormalities related to the reported failure.Failure analysis - unit received with the mayfield 2 lock has up and down movement and the teeth are grinding when rotated.The disk ratchet and retainer was replaced due to wear.General maintenance and replacement of worn parts were performed.Root cause - the reported complaint was confirmed from the evaluation.Evaluation showed that the lock has up and down movement and the teeth are grinding when rotated.The unit needs repair, and replacement of worn parts.General maintenance and cleaning required.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
A facility reported that the rocker arm on mayfield composite series skull clamp (a3059) would not move, it kind of grinds and gets stuck when the clamp was on the patient's skull and the locking knob was unlocked.A 5 minute surgery delay was reported with no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD COMPOSITE SERIES SKULL CLAMP
Type of Device
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12758985
MDR Text Key281745442
Report Number3004608878-2021-00623
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253792
UDI-Public10381780253792
Combination Product (y/n)N
PMA/PMN Number
K142238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA3059
Device Catalogue NumberA3059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2021
Date Manufacturer Received10/13/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown