SMITH & NEPHEW, INC. JRN II BCS LCK FEM IMPLANT IMPACT; PRSTHS, KNEE, PATFEMIB, SEMI-CNSTRND, CMNTD, POLY/MET/POLY
|
Back to Search Results |
|
Model Number 74012812 |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/14/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that, during spd cleaning, the attachment arm of the fem implant impact was noticed cracked.As this was noticed in a non-surgical environment, there was not any patient involved.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
H3, h6: the device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned impactor confirms the cam arm has a crack in it, rendering the device inoperable.The device shows signs of significant wear and use.A review of complaint history did not reveal additional complaints for the listed batch for the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|