HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; IMAGINA GASTROSCOPE Back to Search Results
Model Number EG29-I10
Device Problem Failure to Eject (4010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2021
Event Type  malfunction  
Manufacturer Narrative
This device is not distributed in us so that unique identifier is blank.If additional information becomes available, a supplemental report will be filed with the new information.
 
Event Description
The user found that the watter nozzle could not use normally.This event occurred at the time of during use.There was no report of patient harm.
 
Manufacturer Narrative
This device is not distributed in us so that unique identifier is blank.If additional information becomes available, a supplemental report will be filed with the new information.
 
Event Description
The user found that the watter nozzle could not use normally.This event occurred at the time of during use.There was no report of patient harm.
 
Manufacturer Narrative
Evaluation summary: water/air knob rubber ring was broken.Water/air knob rubber ring was consumable.After soaking in disinfectant for a long time, it would expand, deform or be damaged, resulting in unable to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
IMAGINA GASTROSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key12776125
MDR Text Key286251122
Report Number9610877-2021-01508
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K131902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG29-I10
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/01/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-