The logfile and further information was requested but unfortunately was not made available for the investigation as the device was serviced by a third-party vendor.In general, a reinstall vent alarm indicates a restricted ventilation capability of the ventilator, the possible causes can be manifold.It can be caused by an inspiratory volume error, what means that the tidal volume measured by the inspiratory flow sensor differs at least 25% from the tidal volume applied by the ventilator, a loss of vacuum pressure inside the ventilator piston, e.G.Caused by a faulty diaphragm, a significant leakage in the patient system or the presence of expiratory gas flow during inspiration, e.G.Caused by a sticking valve.A likely root cause is a loose inspiratory elbow nozzle causing a leak between the flow sensor and the breathing system block, which is detected by the device and alarmed immediately.However, in all cases automatic ventilation is restricted.Manual ventilation as well as monitoring functionality is not affected.Based on the available information the exact root cause can¿t be determined.The device alarmed correctly and according to its safety concept for reinstall vent.
|