Model Number FGS-0636 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, the capsule failed to deploy after they broke the handle and pulled the wire.After pulling and twisting the wire, the customer was able to get the capsule to deploy from the delivery device and attach to the esophagus.There was no patient and user harm, and no intervention was required.There was nothing unusual about the patient or the procedure and an endoscopy had been performed prior to the procedure and showed the esophagus to be normal.No lubrication was used to facilitate the placement of the capsule.
|
|