• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII QUICK CONNECT HANDLE; PROSTHESISKNEEFEMOROTIBIALSMICONSTRAINDCMNTDMTALPOLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII QUICK CONNECT HANDLE; PROSTHESISKNEEFEMOROTIBIALSMICONSTRAINDCMNTDMTALPOLYMER Back to Search Results
Model Number 71440044
Device Problems Break (1069); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2021
Event Type  malfunction  
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The ball bearing of the device is jammed and does not move, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.
 
Event Description
A gii quick connect handle was reported to be broken as the device is unable to achieve connection.No case involved, therefore no patient was involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII QUICK CONNECT HANDLE
Type of Device
PROSTHESISKNEEFEMOROTIBIALSMICONSTRAINDCMNTDMTALPOLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
MDR Report Key12780069
MDR Text Key281395622
Report Number1020279-2021-07928
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010213860
UDI-Public03596010213860
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440044
Device Catalogue Number71440044
Device Lot Number07DM00812
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/02/2007
Is the Device Single Use? No
Type of Device Usage A
Patient Sequence Number1
-
-