  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH SLIDING CORE, UHMPWE,6M (STERILE PACKAGED); PROSTHESIS, ANKLE, UNCEMENTED, NON-CONSTRAINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH SLIDING CORE, UHMPWE,6M (STERILE PACKAGED); PROSTHESIS, ANKLE, UNCEMENTED, NON-CONSTRAINED Back to Search Results
Model Number 400-140F
Device Problems Break (1069); Fracture (1260)
Patient Problems Failure of Implant (1924); Implant Pain (4561)
Event Date 01/22/2021
Event Type  Injury  
Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device disposition is unknown.
 
Event Description
A revision surgery was reported to stryker - reason for revision "fractured poly pain"."fractured 6mm poly replaced with a 7mm without difficulty" "no abnormalities, no concerns with the surgical site".
 
Event Description
A revision surgery was reported to stryker - reason for revision "fractured poly pain"."fractured 6mm poly replaced with a 7mm without difficulty" "no abnormalities, no concerns with the surgical site".
 
Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided.The case is about an 80yo female patient with a bmi of 27.6 that was implanted with a star poly in 2016, which broke and provoked pain in 2021 (5 years of implantation).More detailed information about the complaint event such as x-rays, as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIDING CORE, UHMPWE,6M (STERILE PACKAGED)
Type of Device
PROSTHESIS, ANKLE, UNCEMENTED, NON-CONSTRAINED
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
marilyne chaumont
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key12785790
MDR Text Key280573399
Report Number0008031020-2021-00451
Device Sequence Number1
Product Code NTG
UDI-Device Identifier00886385026923
UDI-Public00886385026923
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number400-140F
Device Catalogue Number400140F
Device Lot Number1450195
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age80 YR
Patient SexFemale
Patient Weight77 KG
-
-