KAVO DENTAL TECHNOLOGIES, LLC HELIOS; DENTAL LIGHT Back to Search Results
Model Number HLC
Device Problem Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2021
Event Type  malfunction  
Manufacturer Narrative
Pelton and crane was contacted alleging that a light fell grazing the patient.No injury was reported.A photo of the light was provided by the authorized dealer technician upon request.The dealer technician ordered the required parts to repair the light.Follow up was made with the doctor's office by phone on (b)(6)2021, and the receptionist confirmed that the light had been repaired during the week of (b)(6) 2021 and was functional with no additional problems.This concludes this report.
 
Event Description
An authorized pelton & crane dealer technician called and stated that pelton & crane helios dental light fell down towards the floor grazing the patient while sitting in the chair.There were no serious injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HELIOS
Type of Device
DENTAL LIGHT
Manufacturer (Section D)
KAVO DENTAL TECHNOLOGIES, LLC
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
KAVO DENTAL TECHNOLOGIES, LLC
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
david waters
11727 fruehauf drive
charlotte, NC 28273
7045877297
MDR Report Key12788137
MDR Text Key280597854
Report Number1017522-2021-00007
Device Sequence Number1
Product Code EAZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Distributor
Reporter Occupation Dentist
Type of Report Initial
Report Date 11/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHLC
Device Catalogue Number0.848.5100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/12/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-