• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH PHILIPS INTELLIVUE MICROSTREAM CO2 EXTENSION; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH PHILIPS INTELLIVUE MICROSTREAM CO2 EXTENSION; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number 862393
Device Problems Display or Visual Feedback Problem (1184); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2021
Event Type  malfunction  
Event Description
Etco2 (end-tidal carbon dioxide) module reading co2 of 150 with static waveform.After troubleshooting, the monitor read "etco2 occluded".After replacing the inline etco2, the same message persisted.Respiratory therapist changed the module and the readings returned to baseline.Proving the module to be defective.There was no harm in this event.This is not a recurring issue at our facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHILIPS INTELLIVUE MICROSTREAM CO2 EXTENSION
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
3000 minuteman road
mail stop 4202
andover MA 01810
MDR Report Key12819371
MDR Text Key280815670
Report Number12819371
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number862393
Device Catalogue Number862393
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA11/04/2021
Date Report to Manufacturer11/16/2021
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age730 DA
Patient SexMale
-
-